Journal of Clinical Pediatrics >
Leadless cardiac pacemaker implantation via the internal jugular vein: clinical practice and reflection of a 15kg child with third-degree atrioventricular block after congenital heart surgery
Received date: 2025-04-07
Accepted date: 2025-05-26
Online published: 2025-06-27
Objective To explore the safety and feasibility of implanting leadless pacemakers via the internal jugular vein in children with congenital heart disease (CHD) of small age and low weight who have developed third-degree atrioventricular block (AVB) after surgery. Methods A retrospective analysis was conducted on the clinical data of a child with grade three AVB after congenital heart disease surgery, as well as the implantation of leadless pacemaker through the internal jugular vein. Results A 6.5-year-old girl (weighing 15 kg) with syncope secondary to third-degree AVB following congenital heart defect repair underwent successful implantation of an AVEIR leadless cardiac pacemaker via the internal jugular vein after comprehensive evaluation of vascular diameter and cardiac dimensions. Postoperative pacing parameters were good, clinical symptoms were eliminated, and no pacemaker-related complications occurred. Conclusions In low-weight young children, after thorough preoperative vascular and cardiac function evaluations, the intracervical vein approach for leadless cardiac pacemaker implantation proves a safe and feasible choice.
HE Shuang , XU Xin , ZHOU Xue , LIU Qian , ZHANG Lei , TIAN Jie , LYU Tiewei . Leadless cardiac pacemaker implantation via the internal jugular vein: clinical practice and reflection of a 15kg child with third-degree atrioventricular block after congenital heart surgery[J]. Journal of Clinical Pediatrics, 2025 , 43(7) : 549 -553 . DOI: 10.12372/jcp.2025.25e0359
| [1] | 李小梅, 丁燕生, 马长生, 等. 中国儿童心血管植入性电子器械专家共识[J]. 中国心脏起搏与心电生理杂志, 2023, 37(1): 1-11. |
| Li XM, Ding YS, Ma CS, et al. Chinese expert consensus on pediatric cardiovascular implantable electronic devices[J]. Zhongguo Xinzangqibo Yu Xindianshengli Zazhi, 2023, 37(1): 1-11. | |
| [2] | 陈轶维, 朱荻绮, 杜欣为, 等. AVEIR无导线起搏器治疗儿童病态窦房结综合征1例[J]. 中华儿科杂志, 2025, 63(2): 195-197. |
| Chen YW, Zhu DQ, Du XW, et al. AVEIR leadless pacemaker implantation in a child with sick sinus syndrome[J]. Zhonghua Erke Zazhi, 2025, 63(2): 195-197. | |
| [3] | Writing Committee Members; Shah MJ, Silka MJ, et al. 2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients[J]. Heart Rhythm, 2021, 18(11): 1888-1924. |
| [4] | Tam MTK, Cheng YW, Chan JYS, et al. Aveir VR real-world performance and chronic pacing threshold prediction using mapping and fixation electrical data[J]. Europace, 2024, 26(3): euae051. |
| [5] | Neuzil P, Hubbard C, Doshi RN, et al. Implantation techniques for a helix-fixation dual-chamber leadless pacemaker[J]. Heart Rhythm, 2024, 21(12): 2552-2562. |
| [6] | Rodriguez J, Cortez D. Aveir retrievable, 38-mm length, leadless pacemaker implantation in a 23-kg pediatric patient with congenital heart disease[J]. Pacing Clin Electrophysiol, 2024, 47(3): 398-400. |
| [7] | Wong A, Petersen MJ, Davidson S, et al. First report of an Aveir retrievable leadless pacemaker in a pediatric patient, via internal jugular vein access[J]. Pacing Clin Electrophysiol, 2023, 46(8): 824-826. |
| [8] | Breeman KTN, Tjong FVY, Miller MA, et al. Ten years of leadless cardiac pacing[J]. J Am Coll Cardiol, 2024, 84(21): 2131-2147. |
| [9] | Wang Y, Zhang Y, Yao G, et al. Echocardiographic measurements in normal Chinese adults (EMINCA) II focusing on left ventricular and left atrial size and function by three-dimensional echocardiography[J]. Front Med, 2024, 18(4): 649-663. |
| [10] | 中国医师协会心律学专业委员会,中华医学会心电生理和起搏分会. 无导线起搏器临床应用中国专家共识(2022)[J]. 中华心律失常学杂志, 2022, 26(3): 263-271. |
| Chinese Society of Arrhythmias, Chinese Society of Pacing and Electrophysiology. Chinese expert consensus on leadless pacemaker (2022)[J]. Zhonghua Xinlushichangxue Zazhi, 2022, 26(3): 263-271. | |
| [11] | Saleem-Talib S, van Driel VJ, Nikolic T, et al. The jugular approach for leadless pacing: a novel and safe alternative[J]. Pacing Clin Electrophysiol, 2022, 45(10): 1248-1254. |
| [12] | López álvarez JM, Pérez Quevedo O, Santana Cabrera L, et al. Vascular ultrasound in pediatrics: estimation of depth and diameter of jugular and femoral vessels[J]. J Ultrasound, 2017, 20(4): 285-292. |
| [13] | Molitor N, Saleem-Talib S, Ramanna H, et al. Leadless pacemaker implantation via the internal jugular vein[J]. Europace, 2024, 26(8): euae199. |
| [14] | Wong A, Yeh J, Davidson S, et al. Aveir VR, retrievable leadless pacing in the young[J]. Pacing Clin Electrophysiol, 2024, 47(8): 988-993. |
/
| 〈 |
|
〉 |